Evaluation of the Kuduwave™ audiometer for compliance with standards for hearing conservation purposes. (Research Report)

The technical specifications of the Kuduwave™ 5000 screening audiometer were found to comply with the relevant standards for hearing conservation programme purposes. The audiometrist must ensure compliance with the calibration requirements annually, or three-monthly if used as a mobile device, and that the ambient noise levels in the test room comply with the standards set for the test environment for the Kuduwave™ 5000 headset.

This study was not to validate whether or not the KUDUwave™ complies with Saouth African occupational Health Standards and there foresuitable for Occupational Health Testing. 

 

View the full article here

 